top of page
calibration GUIDE

SONITUS SYSTEMS | EM2030 | NOISE MONITOR


SONITUS SYSTEMS
EM2030
calibration
Tools required
-
Mobile phone with WIFI, Tablet or PC.
-
Tools & Allen Keys
-
Calibrator (94 dB or 114 dB)
​
procedure
-
Remove the windshield from the microphone.
-
Place the calibrator on the end of the microphone.
-
Switch ON the calibrator
​
WITH YOUR device CONNECT TO THE MONITOR VIA WIFI
See next tab...

Accuracy : IEC 61672 Class 1
Dynamic Range : 17 to 121dB(A)
Frequency Range : 20Hz to 20kHz
Frequency Weighting : A and C weighting
Parameters: LEQ, L5, L10, L50, L90, L95, LMAX
Measurement Period: 1, 5, 10, 15 or 30 minutes
Data Storage: 5 years (5-minute logging)
Procedure: Automatic measurement and logging
Power Input: 110V-240V AC / 12V DC
Power Consumption: 8 - 16V DC (battery option) Maximum Current: 2.4W - Peak 4.5W